Back Manifold Plate

Thermo Scientific

To replace the back manifold plate:

Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
127.
Note You do not have to remove the ion source cartridge.
2. Pull the column fully out of the transfer line and move the instrument away from the GC.
3. Slide the top cover panel toward the back of the instrument and lift it off.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Figure 291. Removing the Right Side Panel Screws
Replacing Components of the Manifold
"Shutting Down the TSQ 9000 System"
Right Side
Panel Screws
TSQ 9000 Hardware Manual
3
Advanced Troubleshooting
269