Thermo Scientific TSQ 9000 Hardware Manual page 268

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
240
TSQ 9000 Hardware Manual
4. Disconnect the vacuum interlock cable from the connector.
Figure 245. Disconnecting the Vacuum Interlock Cable
5. Unscrew and then lower the vacuum interlock handle.
6. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 246. Removing the Vacuum Interlock Shield
Springs
Lower the Vacuum Interlock Handle
Vacuum
Interlock
Cable
Connector
Vacuum
Interlock
Handle
Thermo Scientific