Replacing The Lens Heater Block; Reattaching The Lens/Source Heater - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
16. If you need to replace the lens heater block, remove the screw holding the lens plate and
springs to the lens heater block.
Figure 152. Replacing the Lens Heater Block
17. Remove and replace the lens heater block.
18. Attach the lens plate and springs to the lens heater block.
19. Attach the source heater block.
20. Reattach the lens/source heater to the new source heater block and set them next to the
rest of the ion source.
Figure 153. Reattaching the Lens/Source Heater
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Lens Heater Block
Lens Plate
and Spring
Screw
Lens/Source
Heater
Screws and Cone
Springs
TSQ 9000 Hardware Manual
175

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents