Replacing The Electrometer Board - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
356
TSQ 9000 Hardware Manual
6. Remove the four screws holding the electrometer board in place
Figure 382. Replacing the Electrometer Board
Screws
7. Remove and replace the electrometer board if necessary.
8. Reattach the screws that hold the electrometer board in place.
9. Reattach the electrometer shield and screws.
Tip To prevent the screws from falling down into the instrument, place them on the shield
before you reattach it to the chassis.
10. Reconnect the electrometer board cable.
11. Reattach the top cover panel and tighten the screw holding it in place.
12. Restart your system by following the instructions in
"Restarting the System"
on
Thermo Scientific
page
145.