Reassembling the Advanced EI Ion Source Cartridge

Thermo Scientific
Figure 228. Removing the Advanced EI Ion Source for Cleaning
12. Twist the Vespel cap off the ion source cartridge and remove the filament. You may now
disassemble the Advanced EI ion source and clean or replace individual parts. See
Figure
229.
Figure 229. Advanced EI Ion Source Cartridge
The ion source parts are assembled the same way as in
Source Cartridge"
on
page 210
instead of the source holder.
1. Add the filament to the correctly assembled Advanced EI ion source cartridge. See
Figure
230.
"Reassembling the Extractabrite Ion
up until the filament although you may use gloved hands
3
Advanced Troubleshooting
Replacing Components of the Ion Source
TSQ 9000 Hardware Manual
231