Advanced Ei Ion Source - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

4
Upgrade Equipment

Advanced EI Ion Source

Advanced EI Ion Source
420
TSQ 9000 Hardware Manual
If you did not order a TSQ 9000 system with an Advanced EI ion source, you can order a
system later. You must change the source magnets on the manifold door to use an Advanced
EI source. To install an Advanced EI ion source.
1. Follow the steps in
"Ion Source Cartridge"
source cartridge.
2. Follow the steps in
"Powering Off the TSQ 9000 System"
TSQ 9000 system.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
4. Open the manifold door.
Figure 445. Opening the Manifold Door
on
page 197
to remove the Extractabrite ion
on
page 10
Screws
to shut down the