7. Remove the controller interface board from the instrument.
8. Look through the top of the instrument and locate both RF boards. The front RF board
is the second board standing on end toward the center of the instrument. The rear RF
board is the third board standing on end toward the center of the TSQ 9000 instrument.
Use a T20 Torxhead screwdriver to remove the screw holding the board and bracket to
the chassis.
Figure 394. Locating the RF Boards and Brackets
.
9. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument. Lift the RF boards out of the instrument and set them aside.
Rear RF
Front RF
Board
Board
Brackets