20. Slide the source holder, which is in the TSQ 9000 Toolkit, onto the end of the ion source
cartridge. The opening of the source holder is designed to accommodate the handles of
the ion volume.
Figure 198. Attaching the Source Holder
Wing of Ion
Volume
Flat surface lines
up with Filament
Hole
Source Holder
21. Twist the holder to disengage the ion source cartridge from the tool.
Note The repeller locking nut is now loose on the repeller spring. Be careful not to
tip the source or the components will fall out.
22. Set the ion source cartridge and holder on a clean surface.
Figure 199. Removing the Ion Source Cartridge
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Filament Hole
Ion Volume
Handles Fit Here
TSQ 9000 Hardware Manual
209