Advanced Troubleshooting
Replacing Components of the Ion Source
214
TSQ 9000 Hardware Manual
9. Insert the large flat end of the repeller so that it rests on top of the ion volume.
Figure 206. Inserting the Repeller into the Source Sleeve
10. Slide the repeller spring onto the repeller.
Figure 207. Inserting the Repeller Spring into the Sleeve
Large Flat
End of
Repeller
Repeller
Spring