Figure 96. Identifying the Set Screw on the Source Heather Block
14. Remove this screw with a T10 screwdriver. It is important to remove this screw
completely so that it cannot damage the transfer line during disassembly/reassembly.
Place this screw in a clean place until it is needed for reassembly. See
Figure 97. Removing the Set Screw
Source Interface
Board
Replacing Components of the Analyzer
Set
Screw
TSQ 9000 Hardware Manual
3
Advanced Troubleshooting
Set Screw
Figure
97.
Calibration and
Reagent Gas
Tubes
Right Screw
133