Thermo Scientific TSQ 9000 Hardware Manual page 239

Table of Contents

Advertisement

Thermo Scientific
2. Insert the ion cartridge sleeve into the source holder.
Figure 201. Inserting the Sleeve into the Source Holder
3. Align the long tooth of lens 3/RF lens with the notch on the bottom of the sleeve and
drop the lens into the sleeve.
Figure 202. Inserting Lens 3/RF Lens into the Source Sleeve
3
Replacing Components of the Ion Source
Long Tooth on Lens 3/RF Lens
Tooth Fits Here
TSQ 9000 Hardware Manual
Advanced Troubleshooting
211