Removing The Power Module Screws - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
4. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
Figure 386. Removing the Power Module Screws
Screws
5. Make sure the electrometer board cable is not in the way when you remove the power
module.
6. Grasp the power supply system and lift it out of the instrument. It fits snugly, so you may
have to gently shake it as you lift it out.
3
Advanced Troubleshooting
Replacing the Power Supplies
TSQ 9000 Hardware Manual