3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
308
TSQ 9000 Hardware Manual
8. Use a 2.5 mm nut driver to remove the standoffs on the rod driver board.
Figure 328. Disconnecting the Rod Driver Board from the RF Board
Screws
9. Incrementally, remove the 40-pin connector on the rod driver board from the front RF
board. and the 7-pin connector on the rod driver board away from the front RF board.
Figure 329. Disconnecting the Rod Driver Board 40-Pin Connector
Standoffs
Rod Driver Board
40-Pin Connector
Thermo Scientific