Advanced Troubleshooting
Replacing the Calibration Gas Components
374
TSQ 9000 Hardware Manual
Note This screw is required for good ground contact.
26. Reinstall the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 398. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
27. Push the board down into the connectors.
28. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
reattach the screw holding the lens driver board in place on the chassis.