Attaching The O-Ring To The Ion Gauge Tube; Installing The Ion Gauge - Thermo Scientific TSQ 9000 Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
11. Place the o-ring from the plug on top of the tube.
Figure 473. Attaching the O-Ring to the Ion Gauge Tube
O-Ring
12. Use a T20 Torxhead screwdriver to attach the ion gauge to the instrument.
Figure 474. Installing the Ion Gauge
13. Replace the right side panel and attach the screws holding it in place.
14. Reattach the top cover panel and tighten the screw holding the top cover panel in place.
15. Restart your system by following the instructions in
4
Upgrade Equipment
Ion Gauge and Tube Shield
Tube
"Restarting the System"
TSQ 9000 Hardware Manual
on
page
145.
439

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents