Installing A Heated Sample Transfer Line - Thermo Scientific TRACE 1300 Instruction Manual

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

3
Components Installation

Installing a Heated Sample Transfer Line

Installing a Heated Sample Transfer Line
94
TRACE 1310 Auxiliary Oven Instruction Manual
This section provides the instructions for installing a heated sample transfer line into the main
oven.
To install a heated sample transfer line
1. Remove the partial cut shaped plate marked Heated Sample Transfer Line from the back
panel of the instrument. Carefully insert the transfer line and its insulator tube into the
duct carried out. See
Figure 137. Installing a Heated Sample Transfer Line (1)
2. Push the transfer line and its insulator tube into the rear of the instrument up to reach the
main oven. See
Figure
Figure 138. Installing a Heated Sample Transfer Line (2)
Figure
137.
Partial Cut Shaped Plate
Transfer Line with Insulating Tube
138.
Thermo Scientific