Correcting leaks

1.
Use an electronic leak detector to check all areas of the inlet and plumbing
that are potential sources of a leak.
2.
Tighten loose connections to correct leaks, if necessary. You may need
to repeat the leak test.
3.
If the pressure drop is now 1.0 psi or less, you can consider the inlet
system leak-free. If the pressure drop is much greater than this, continue
to search for leaks and repeat the pressure test.
18 of 20
EPC module serviceable parts
Inlets
Agilent 6890 Gas Chromatograph Service Manual