Installing A Type 2 Manifold - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Installing a Type 2 manifold

WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off all detector gases at their supply, then turn off the main power
switch and unplug the power cord.
Caution
Always hold the manifold as shown below to avoid damaging board
components.
1.
While holding the manifold by the black plastic frame as shown below,
slip the label through the slot in the mounting bracket, and align the
bracket holes over the gas fittings.
Hold manifold by
the black plastic frame.
Slip label tag through
the slot in the bracket.
2.
Finger-tighten three 7/16-inch hex nuts over the fittings to hold the
bracket in place. It is very important that you do not tighten the nuts yet.
Jun 2001
Micro-cell Electron Capture Detector (µ-ECD)
(FPD shown)
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
341
13 of 26

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents