Typical values

3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
Zero
Att 2^
Cht sp
Ar Rej
Sensitivity
Discrimination C
3396B or 3396C or 3397A integrator sample conditions with INET input
[6890 INET output, Range 0]
>>Set DATE and TIME<<
Zero
Att 2^
Cht sp
Ar Rej
Sensitivity
Discrimination C
Agilent ChemStation
Sensitivity
Discrimination C
Agilent Chemstation—ASTM noise
Measured with 6890 isothermal @ 100°C
10 minute blank run, noise range(s) >1 minute
Performance + noise report
TCD ASTM noise ≤ 0.05733 display units (25 µV/display unit)
Detector signal set to 5 Hertz
Jun 2001
TCD checkout conditions and chromatogram
10
Pk wd
7
Thresh
1
Stop time
1000
C
, C
, C
area counts
14
15
16
/C
area ratio
14
16
10
Pk wd
7
Thresh
1
Stop time
100
C
, C
, C
area counts
14
15
16
/C
area ratio
14
16
C
, C
, C
area counts
14
15
16
/C
area ratio
14
16
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms
0.04
5
8.75
SS/COC/PTV PPIP
585,000
515,000
1.00±0.10
1.00±0.10
0.04
5
8.75
SS/COC/PTV PPIP
23,000
20,130
1.00±0.10
1.00±0.10
SS/COC/PTV PPIP
73
65
1.00±0.10
1.00±0.10
610