540
Instrument Specifications
Column oven
Column oven
•
Dimensions: 28 × 31 × 16 cm
•
Operating temperature: 4° C above ambient to 450° C
–
–
•
Temperature setpoint resolution: 1° C
•
Maximum setpoint temperature rate: 120° C/min
•
Actual programming rate: see Figure 540-1
•
Cool-down rates: see Figure 540-2. Rates are greater when using the
optional oven exhaust deflector or cryogenic cooling.
•
Maximum run time: 999.99 min
•
Programming ramps/plateaus: 6/7
•
Ambient rejection: <0.01° C per 1° C
•
Column bleed compensation standard for two channels
Temperature (°C)
500
240 Volts
450
400
350
300
250
200
150
100
50
0
0
100
200
300
400
Time (sec)
Figure 540-1
Typical oven heatup rates,
6890 Series GC
2 of 6
with LN
cryo: –80° C to 450° C
2
with CO
cryo: –55° C to 450° C
2
120 Volts
500
600
700
800
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
Temperature (°C)
500
450
400
350
300
250
200
150
100
50
0
0
100
200
300
400
Time (sec)
Figure 540-2
Typical oven cool-down
rates, 6890 Series GC
500
600
700
800
Jun 2001