Detector
Temperature
Hydrogen flow
Air flow
Const. Makeup Flow, N
Offset, flow off (O–fa)
Offset, flame on (O+fb) <[(O–fa) + 85 display units]
Oven
Initial temp
Initial time
Rate 1
Final temp
Final time
Rate 2
Final temp 2
Final time 1
Jun 2001
FPD checkout conditions and chromatogram
200° C
75±2 mL/min
100±2 mL/min
60±2 mL/min
2
Should be <40 display units
60° C
0 min
25° C/min
110° C
0 min
10°C/min
170°C
3 min
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms