Agilent Technologies 6890 Series Service Manual page 384

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Caution
Overtightening may cause hazardous leaks. Tighten the hex nuts exactly as
described below.
12. Use a wrench to tighten the 7/16-inch hex nuts 3/4 turn only.
13. Using a pair of needle-nosed pliers, remove the top rear panel cutout for
Auxiliary. Also remove any cutouts needed to access other manifolds or
accessories installed in the GC. See Figure 360-8.
Front detector
Auxiliary
Left most screw slot
Back detector
Figure 360-8
14. Place the new top rear panel on its left-most mounting screw. Use the
screw as a hinge and angle the panel while sliding each manifold ID tag
through its cutout in the panel, working from left to right. When all the
tags are through the panel, finish installing the panel on the GC.
Jun 2001
Top rear panel cutouts
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Auxiliary EPC
Replacement procedures
Insert tip of
pliers here
Back inlet
Front inlet
360
11 of 16