Replacement Procedures; Replacing The Ptv Inlet, Pneumatics, And Manifold - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Replacement procedures

In addition to the PTV consumables (inlet adapters, columns, Teflon ferrules,
and septa), the replacement parts in the PTV inlet assembly are:
The entire PTV inlet, pneumatics, and manifold assembly
The PTV manifold assembly
PTV gang weldment and front trap assemblies
The filter
The head assembly (septum or septumless)
The cryo shroud on the inlet body (CO
The PTV thermocouple PCB
The O-rings and restrictors in the gang fitting assembly.
Replacing the PTV
The entire PTV assembly can be easily replaced.
1.
Turn off the oven and the inlet and allow them to cool.
WARNING
Turn off the oven and the inlet and allow them to cool. Turn off all flows at
the initial gas supply. Then turn off the main power switch and unplug the
power cord.
2.
Remove the top cover, the inlet fan cover, the inlet cover, the left side
panel, the top rear panel, the pneumatics chassis cover, and the RFI cover.
3.
Remove the top mounting screw in the PTV manifold assembly.
Jun 2001
Programmed Temperature Vaporization Inlet (PTV)
inlet, pneumatics, and manifold
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
or liquid nitrogen)
2
240
13 of 46

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents