350
Flame Photometric Detector (FPD)
Replacement procedures
Tighten the gang fitting screw firmly to compress the O-rings.
6.
Route the ribbon cable behind the manifold assembly as shown in
Figure 350-16. Then, slide the manifold and bracket assembly into the
slot until the bracket seats flush against the end of the rails. See
Figure 350-17.
Figure 350-16
28 of 36
Be sure the left tube clears the inner edge of the bracket. See
Figure 350-17.
Be sure that the O-rings are in place.
Back view of manifold
Routing the ribbon cable
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Manifold installed with cable routed to left
Jun 2001