Advertisement

Typical values

3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
Noise ([N1])
Measured at Attn. 0, Chart speed 1
Signal
Zero
Att 2^
Cht sp
Ar Rej
Area, Lindane peak
3396B or 3396C or 3397A integrator sample conditions with INET input
[6890 INET output, Range 0]
>>Set DATE and TIME<<
Noise ([N2])
Measured at Attn. (–5), Chart speed 1 [N2] 11.25 mm for 1 min.
measurement
Signal
Zero
Att 2^
Cht sp
Ar Rej
Area, Lindane peak
Agilent ChemStation
Noise ([N3])
Measured by ChemStation, ASTM noise [N3] <0.05733 display units
(5 Hz/display unit)
Signal
Area, Lindane peak
Jun 2001
ECD checkout conditions and chromatogram
10
Pk wd
7
Thresh
1
[Stop time
10
Pk wd
7
Thresh
1
[Stop time
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms
[N1] <9 mm for 1 min. measurement
0.04
7
11.0]
84150× [N1]
0.04
7
12.0]
2627 × [N2]
1650 × [N3]
610
17 of 26