RS-232
RS-232 Standard wiring
Transmitters
Receivers
Jun 2001
External Connectors, 6890A and 6890 Plus
9
5
6
1
5
9
6
1
3 TxD
7 RTS
4 DTR
2 RxD
8 CTS
6 DSR
1 DCD
5 GND
Electrical
Agilent 6890 Gas Chromatograph Service Manual
JP1 (Host/Modem), JP2 (Sampler)
RS-232
Pin Function
1
DCD
INPUT
2
RxD
3
TxD
OUTPUT
4
DTR
OUTPUT
5
GND
6
DSR
INPUT
7
RTS
8
CTS
INPUT
9
unused
(No Connect)
RS-232 Loop back wiring
1
2
3
4
5
6
7
8
9
1210
RS-232
In/Out
INPUT
OUTPUT
INPUT
7 of 14