Warning 113—Aux 3 config changed, method defaulted
Warning 114—Aux 4 config changed, method defaulted
Warning 115—Aux 5 configuration changed, method defaulted
Status message: Aux 3 method changed or Aux 4 method changed or Aux 5
method changed
This message appears when the 6890 GC is turned on and one of the auxiliary
pressure channels has changed. Either the type, sensor range, length,
diameter or something else has changed.
Default setpoints for the aux module will be loaded.
Warning 116—Run log full
Status message: None
The run log can hold 50 lines of information. When it reaches the maximum,
no more information can be stored in the log and this warning will appear.
Remove some information from the run log.
Warning 117—F inl calib deleted
Warning 118—B inl calib deleted
Status message: F inl calib deleted or B inl calib deleted
If you were entering a new inlet calibration for the front or the back inlet and
the 6890 GC crashed or encountered a power failure before the recalibration
was complete, this warning will appear when the instrument is turned on
again.
The calibration for the module will return to the default.
Jun 2001
Warning 113—Aux 3 config changed, method defaulted
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes
830