Advertisement

5.
If the tubing from the gang fitting bends to the left, reshape it so that it
bends up and back from the block as shown in Figure 360-6.
6.
Insert the gang fitting through the cutout in the manifold bracket and
install it onto the new manifold assembly so that the tubing runs back
and away from the fitting.
Tighten the screw on the gang fitting until the gang fitting touches the
manifold.
7.
Route the ribbon cable to the right side of the manifold assembly as shown
in Figure 360-5. Then, slide the manifold and bracket assembly into the
Aux slot until the bracket seats flush against the end of the rails. See
Figure 360-6.
Figure 360-5
Jun 2001
Be sure that the three O-rings are in place.
Be sure the left tube clears the inner edge of the bracket. See
Figure 360-6.
Routing the ribbon cable
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Auxiliary EPC
Replacement procedures
Ribbon cable
360