Advertisement

5.
Insert the gang fitting through the cutout in the manifold bracket and
install it onto the new manifold assembly so that the tubing runs back
and away from the fitting.
Tighten the gang fitting screw firmly to compress the O-rings.
6.
Route the ribbon cable behind the manifold assembly as shown in
Figure 341-11. Then, slide the manifold and bracket assembly into the
slot until the bracket seats flush against the end of the rails. See
Figure 341-12.
Back view of manifold
Figure 341-11
Jun 2001
Micro-cell Electron Capture Detector (µ-ECD)
Be sure the left tube clears the inner edge of the bracket. See
Figure 341-12.
Be sure that the O-rings are in place.
Routing the ribbon cable
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
Made in U. S. A.
Manifold installed with cable routed to left
341