O-ring
Figure 350-13
6.
Remove the ignitor cable.
7.
Unscrew the connector and ignitor assembly from the detector module.
8.
Remove the O-ring, ignitor spacer, and glow plug from the ignitor
assembly. Replace the glow plug. Inspect the O-ring and spacer, and
replace if needed.
9.
Reassemble in reverse order.
Jun 2001
Spacer
Glow plug
Connector
Ignitor cable
Replacing the ignitor glow plug