210
Split/Splitless Inlet
Leak testing the EPC module only
6.
Connect one end of the leak test ribbon cable (G1530-61370) to the ribbon
cable connector on the EPC module and connect the other end to the
appropriate connector on the EPC board. Reconnect the gas supply and
turn on the main power switch. Set the pressure to 25 psi and the flow
to 60 mL/min.
7.
Lay the EPC module on the lab bench and use an electronic leak detector
to locate the leaky component on the module.
To gas supply
Figure 210-16
8.
If the leaky component is serviceable, such as a vent/inlet fitting (see
diagram of serviceable parts, Figure 210-17 ), replace it. Otherwise,
replace the EPC module.
28 of 30
Leak testing the EPC module on the lab bench
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Ribbon cable from