Replacing The Collector Assembly - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

4.
Remove the thermal strap and weldment from the detector pallet.
Disconnect the strap from the weldment (two M7 hex nuts).
Detector pallet
Hex nut
Thermal strap
Detector weldment
M7 hex nuts
Insulation
Figure 310-8

Replacing the collector assembly

WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off any detector gases at their supply. Then turn off the main
power switch and unplug the power cord.
1.
Remove the right side cover of the instrument.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
Jun 2001
Removing the detector block
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
Replacement procedures
Heater/sensor assembly
310
9 of 24