Agilent Technologies 6890 Series Service Manual page 82

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

5.
Note the "Actual" reading on the display and monitor the pressure for
10 minutes.
You can use the stopwatch feature of the 6890 GC to monitor the time.
Press
the time and the pressure reading with the
Inlet
6.
When the system is considered leak tight, the caps may be removed, the
column reinstalled, its dimensions configured at keyboard, and the
desired pressure or flow rate set.
Jun 2001
Time] and then [Enter
[
/
Back Inlet
keys.
]
[
]
If there is less than 0.7 psi pressure loss (0.07 psi/min.), consider
the system leak tight.
If pressure loss is much greater than 0.7 psi, there is a leak that must
be found and corrected. See Correcting leaks later in this section.
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed Inlet
Leak testing—EPC and manual inlets
to start timing, then toggle between
]
Time
and the [Front
[
]
220
13 of 20