341
Micro-cell Electron Capture Detector (µ-ECD)
Replacement procedures
6.
Remove the Torx T-20 screw holding the output pneumatics block(s) to
the manifold and remove the block.
Caution
Always hold the Type 2 manifold by its support bracket to avoid damaging
board components.
7.
Slide the manifold from its slot.
There are two pneumatics blocks on the EPC version of the µ-ECD pneumatics
Note
manifold. The outside block is the anode purge line and the inside block is
the makeup gas line.
Figure 341-9
12 of 26
Removing an EPC detector flow manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Type 2 manifold
Type 1 manifold
Jun 2001