Cool On-Column Inlet
230
Leak testing EPC and manual inlets
Manual
Septum purge vent
EPC
Septum purge
Figure 230-8
Capping the bottom of the inlet and septum purge vent
Note
Make sure that the carrier gas source pressure is at least 35 psi. Carrier source
pressure should always be at least 10 psi greater than the desired inlet
pressure.
11 of 20
Inlets
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual