Agilent Technologies 6890 Series Service Manual page 639

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1120
Valve Box
Installing the valve box
2.
Screw the aluminum standoffs into the valve box bottom plate using one
Torx T-20 screw for each standoff.
Figure 1120-2
3.
Screw the valve box bottom plate onto the top of the GC using three Torx
T-20 screws.
Figure 1120-3
2 of 10
Attaching the standoffs to the bottom plate
Attaching the bottom plate to the 6890 GC
Valves
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001