Figure 350-14
Caution
Always hold the Type 2 manifold by its support bracket to avoid damaging
board components.
7.
Remove the manifold from its slot.
Installing a Type 2 manifold
WARNING
Hydrogen gas is flammable and potentially explosive. Before replacing the
manifold, turn off the hydrogen gas at the source.
Jun 2001
Type 2 manifold
Type 1 manifold
Removing an FPD flow manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Photometric Detector (FPD)
Replacement procedures
Remove the screw
350
25 of 36