Thermal Conductivity Detector (TCD)
330
Theory of operation

EPC detector

The EPC TCD has one flow line for the makeup/reference gas. After an initial
filter frit, it splits into two paths. Each line has a proportional valve controlled
by a pressure sensor, and a non-adjustable restrictor frit. In addition, the
switching gas line has a switching valve that switches paths through the TCD
cell several times a second. The switching line oscillates between two
flowpaths, one for each channel of the TCD cell. The makeup gas line sweeps
past the end of the column and carries the column effluent into the TCD cell.
Vent
Makeup and
reference gas in
Filter
frit
Reference
switching
Proportional
Pressure
valve
valves
sensors
Restrictors
PS
Makeup flow
PS
Pressure
control loops
Figure 330-4
EPC TCD flow diagram

Manually controlled detector

The manually controlled TCD has a pneumatic supply line for the switching
gas (also called modulating or reference gas) flow and one for the makeup
gas. Both of the lines have pressure regulators, an on/off solenoid valve and
a non-adjustable restrictor frit. In addition, the switching gas line has a
5 of 24
Detectors
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual