Using Liquid Nitrogen - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

520
Gases
Cryogenic cooling requirements

Using liquid nitrogen

WARNING
Liquid nitrogen is a hazard because of the extremely low temperatures and
high pressures that may occur in improperly designed supply systems.
Liquid nitrogen can present an asphyxiant hazard if vaporizing nitrogen
displaces oxygen in the air. Consult local suppliers for safety precautions and
design information.
Liquid nitrogen is supplied in insulated Dewar tanks. The correct type for
cooling purposes is a low-pressure Dewar equipped with a dip tube—to deliver
liquid rather than gas—and a safety relief valve to prevent pressure build-up.
The relief valve is set by the supplier at 20 to 25 psi.
WARNING
If liquid nitrogen is trapped between a closed tank valve and the cryo valve
on the GC, tremendous pressure will develop and may cause an explosion.
For this reason, keep the delivery valve on the tank open so that the entire
system is protected by the pressure relief valve.
To move or replace a tank, close the delivery valve and carefully disconnect
the line at either end to let residual nitrogen escape.
Additional requirements for the liquid N
• Nitrogen must be provided to the GC as a liquid at 20 to 30 psi.
• The supply tubing for liquid N
refrigeration and air-conditioning lines is suitable for insulation. Since
pressures are low, insulated copper tubing is adequate.
• The liquid nitrogen tank should be close (only 5 to 10 feet) to the GC to
insure that liquid, not gas, is supplied to the inlet.
12 of 14
must be insulated. Foam tubing used for
2
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
system include:
2
Jun 2001

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents