Agilent Technologies 6890 Series Service Manual page 193

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

260
Solvent Vapor Exit Accessory
Leak testing
1.
Replace the septum of the Cool On-column inlet.
2.
Using a 1/16-inch (SS) union and the appropriate graphite/Vespel
ferrules, connect the detector end (250 µm) of the tri-column assembly
and the transfer line (320 µm) together.
3.
From the 6890 GC keyboard, set the inlet pressure to 50 psi and wait
approximately 15 seconds for equilibration.
Note
If pressure can not be reached, either a very large leak is present or the supply
pressure is not high enough (at least 10 psi greater than desired pressure).
4.
Cap the septum purge vent of the Cool On-column flow module using a
solid (no- hole) 1/8-inch Vespel ferrule (part no. 0100-1372) and a 1/8-
inch Swagelok nut (part no. 5180-4103). As an alternative a 1/8-inch
Swagelok cap can be used.
5.
From the GC keyboard, turn the inlet pressure "OFF".
6.
Note the "Actual" reading on the display and monitor for 10 minutes. You
can use the stopwatch feature of the 6890 GC to monitor the time by
pressing [Time] and then [Enter] to start timing.
7.
If there is less than 1.5 psi pressure loss, consider the system to be leak
tight.
8.
If pressure loss is much greater than 1.5 psi, there is a leak that must be
found and corrected.
8 of 10
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001