Instrument Specifications

Detectors

Detectors
All detectors are available with electronic control of gases.
All detectors except the µ-ECD are available with manual control of gases.
All detectors (manual or electronic) include electronic on/off of all gases.
Detectors available:

FID

450° C maximum operating temperature
Automatic flame ignition from the keypad or ChemStation
Flame out detection
MDL: <5 pg carbon/s as propane using nitrogen carrier and 0.29 mm jet
Linear dynamic range: 10

NPD

400° C maximum operating temperature
Automatic baseline adjusting via keypad or ChemStation
MDL: <0.4 pg N/s, <0.2 pg P/s with azobenzene/malathion mixture
4 of 6
Flame ionization detector (FID) for packed or capillary columns
Flame ionization detector (FID) optimized for capillary columns
Thermal conductivity detector (TCD)
Electron capture detector (ECD)
Micro-cell electron capture detector (µ-ECD)
Nitrogen-phosphorus detector (NPD) for packed or capillary
columns
Nitrogen-phosphorus detector (NPD) optimized for capillary
columns
Flame photometric detector (FPD), single or dual wavelength
Atomic emission detector (AED)
Mass selective detector (MSD)
7
Site Preparation
Agilent 6890 Gas Chromatograph Service Manual
(± 10%) with nitrogen carrier and 0.29 mm jet