360
Auxiliary EPC
Replacement procedures
Mounting bracket
Gang fitting
installs here
Apply blank label
Ribbon cable
Figure 360-4
3.
Peel the blank label from its backing and paste it on the mounting bracket
over the screw heads. See Figure 360-4.
4.
If you need to change one or more auxiliary channel frits, do so now. See
the Agilent 6890 GC Operating Manual/CD-ROM.
8 of 16
(FPD shown above)
Bracket mounted onto the manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Finger-tighten nuts.
Do not use a wrench.
7/16-inch hex nuts
Label ID tag
through slot
in bracket
Jun 2001