Agilent Technologies 6890 Series Service Manual page 437

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

430
Power/Electronics Replacement
Replacing the main board
Beeper
P11
P1
GND
P12
Oven Access
GND
Cutout
Cutout
P13
P17
3V
Lithium
battery
4 ROM
Sockets
Figure 430-5
9.
Holding the main board by the metal bracket at the top left and the
capacitors at the bottom right, slide it slightly to the left and pull it out
of the instrument being careful not to catch the board on any wiring
harnesses.
10. Installation of the new board is the reverse of this procedure.
Caution
Changing the board loses the GC serial number.
6 of 26
P16
P2
DSP
Gate
Array
CPU
+15V
–15V
+5v
Gate
Array
Screw locations on the main board (6890A/Plus board shown.
6890N is the same.)
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
P21
P22
P15
J8
F4 F3
F2 F1
Hole for
transformer
cable
Capacitors
Capacitor
P3
J1
J2
JP1
JP2
J4
P18
J6
–24V
J5
+24V
Jun 2001