Manually Controlled Inlet; Split Mode - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Manually controlled inlet

Split mode

For a manual inlet in split mode, V1 is a manually regulated mass flow
controller that controls total inlet flow. PCV is a three-way purge control valve
that remains in the normally open position to allow flow out of the split vent.
V2 is a manually controlled back pressure regulator in the split vent path
controlling the inlet pressure and thus, flow through the column. SPR is a
manually controlled septum purge regulator used to control the septum purge
flow. This flow is set to approximately 3.0 mL/minute for operation.
Total Flow
Control
104 mL/min
V1
Figure 210-3
Jun 2001
Septum holder
PCV
Purge Control Valve
Manual control split mode flow diagram. 100:1 split ratio
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet
Manually controlled inlet
Septum purge
regulator
Septum purge
SPR
3 mL/min
Inlet pressure
control
Inlet Vent Flow (split)
100 mL/min
V2
To Detector
1 mL/min Column Flow
210
5 of 30