410
Covers and Fans
Removing the pneumatics area fan
Air flow
Figure 410-18
3.
Remove the Torx T-20 screw from the bracket and slide the bracket out
of the back of the instrument
4.
Lift the retaining clip securing the fan in the bracket and slide the fan
out of the two retaining tabs. Remove the old fan.
5.
Lift the retaining tab and insert the new fan. Be sure to orient the fan so
that the arrow points downward when installed and the wires are on the
right side. Slide the fan under the retaining tabs until it snaps into place.
6.
Slide the tabs on the back of the fan bracket into the corresponding slots
underneath the EPC board area and reinstall the screw.