Agilent Technologies 6890 Series Service Manual page 721

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1230
Power Supply
Replaceable AC board fuses
Replaceable AC board fuses
There are four fuses on the AC power board.
The two glass fuses are identical for all power options.
The two ceramic fuses come in two types: one for the 120 V power option
and another type for all other power options.
When replacing both the glass and ceramic fuse types, always replace them
in pairs.
Table 1230-4
I.D.
Description
F1
Ceramic/Type F
F2
Ceramic/Type F
F3
Glass/Type F
F4
Glass/Type F
4 of 10
Replaceable Fuses on the AC Power Board
System
Power rating
Oven heater
120 V systems: 20A/250 V
200 V–240 V systems: 15A/250 V
Oven heater
120 V systems: 20A/250 V
200 V–240V systems: 15A/250 V
All other systems
8A/250 V
except heater
All other systems
8A/250 V
except heater
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Part no.
2110-0098
2110-0054
2110-0098
2110-0054
2110-0036
2110-0036
Jun 2001

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents