350
Flame Photometric Detector (FPD)
Replacement procedures
Figure 350-8
9.
Remove the transfer line nut and its ferrule, the heater/sensor cable
assembly, and the heated block.
Figure 350-9
16 of 36
Ignitor cable
Transfer line nut
Heater/sensor
cable assembly
Heater block
U-clamp screws
Replacing the transfer line fused silica liner
Liner
Transfer tube
Removing the transfer tube
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001