Agilent Technologies 6890 Series Service Manual page 347

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

350
Flame Photometric Detector (FPD)
Replacement procedures
6.
Remove the four screws in the PMT adapter flange and remove the flange.
Use care as a quartz window is exposed and may fall out. Clean the
window using lens tissue.
Clamp
Figure 350-5
10 of 36
FPD disassembly
Detectors
Agilent 6890 Gas Chromatograph Service Manual
0-ring
Window, second heat shield
0-ring
Flange adapter
Flange ring
O-ring,
Viton
Jun 2001