Figure 230-6
4.
Remove the 1/8-inch Swagelok fitting for the gas supply.
5.
Remove the plumbing block from the front of the manifold (one Torx T-10
screw).
6.
Remove the long screw (Torx T-20) from the top of the manifold and slide
the manifold out of the back of the GC.
7.
Reinstallation is the reverse of removal.

Replacing the supply fitting on a Type 1 EPC flow manifold

See Split/Splitless Inlet for the detailed procedure.
Jun 2001
Removing the EPC flow manifold
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Cool On-Column Inlet
Replacement procedures
230