Remove detector
top cover
Remove anode signal wire
Figure 341-15
4.
Remove the screw and clamp on the electrical interconnect.
5.
Remove one Torx T-20 screw from each end of the signal board. (Do not
remove the screw on the top of the cover.)
Jun 2001
Micro-cell Electron Capture Detector (µ-ECD)
Removing the top cover and anode signal wire
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
341
19 of 26